PMX PU INLET / SCRUBBER KFG

  • Standard reference sampling system for PAHs/BaP according to Directive 2004/107/EC and CEN EN 15549: PM10 pre-impactor (CEN EN 12341) with attached ozone scrubber followed by PU cartouche (with pre-filter)
  • Also available with back-up filter without PU cartouche
  • Use of PM2,5 and PM1,0 pre-impactors possible
  • Flow rate 2,3 operating -m³/h (also 1,0m³/h possible)
  • Filter cassettes for filters with 47 mm and
    50 mm diameters
  • 2 – 3 PU foam pieces of 45 mm diameter and 50 mm height
  • Operating time of the scrubber: approx. 1 month
  • Refreshing of the scubber at 150°C over approx. 8 h